SKCK SKCKHui Ma · Flow measurement & instrumentation

Home / Articles / Not enough straight run?

GB/T 2624.1 · JB/T 7252

Not enough straight run?

Updated 2026-11-21 · about 4 minutes

"There is not that much straight pipe" is the most common site complaint. But a short straight run does not mean the measurement fails — it means you pay for it in uncertainty, and there are three cheaper ways out.

What the A and B columns really mean

GB/T 2624.1 splits the upstream straight run into an A column and a B column, based on the number and type of upstream fittings:

  • A-column length — when met, the standard predicts the flow with no additional uncertainty;
  • B-column length — roughly two thirds of the A column; when met, add 0.5% to the uncertainty budget;
  • shorter than the B column — this part of the standard can no longer predict the uncertainty; use a flow conditioner or a real-flow calibration.

Note the wording: a short straight run does not make the data unusable, it makes the data usable at the price of 0.5%. In custody-transfer projects that 0.5% is often the whole margin available.

The downstream requirement is usually shorter — often 3 to 5 diameters, depending on the device and the tapping arrangement — because the flow re-develops progressively there.

Three ways out

  1. Accept the 0.5%. Cheapest, fine for process control. Just state it in the technical agreement and the meter data sheet.
  2. Fit a flow conditioner. A straightener or honeycomb conditioner in front of the plate buys a stable velocity profile at the cost of pressure loss. The conditioner itself has to pass the compatibility test in the standard — buying one off the shelf and fitting it is not enough.
  3. Change the device or the concept. A Venturi tube needs far less straight run than an orifice plate; a wedge device can carry an internal conditioner down to 0.5D; and the integral orifice flow meter puts plate, tap and conditioner into one valve section — the engineering compromise when space is the constraint.

Choosing a flow conditioner

The key parameter is the resistance coefficient K:

ConditionerKCharacter
Straightener≈ 0.25low loss, cheap, needs length
19-tube bundle straightener (1998)≈ 0.75 specified in GB/T 2624.2; integrated in the integral orifice
Zanker type≈ 5best performance, highest loss

Source: GB/T 2624.1 on flow conditioners.

The logic is straightforward: if the pressure loss budget allows and there is room, use the cheap straightener; if loss is tight but accuracy is critical, use the Zanker; if there is no room at all, use the integral orifice flow meter.

The integral orifice flow meter

JB/T 7252-2018 integrates the orifice plate, the tap body and the 19-tube bundle straightener into a single valve section, trading space for straight run. Its boundaries:

  • flange taps only;
  • PN16–PN250 / Class150–Class1500;
  • DN50–DN700 (NPS2–NPS28);
  • −29 °C to 121 °C;
  • three constructions: type G (dual chamber, plate changed online), type P (single chamber, liftable) and type J (fixed).

The trade-off is clear: if you need to change the plate online (range adjustment), take type G; if you only want to save straight run, type J is enough and the price difference is significant. Also remember that the body flange can stand proud into the tapping position — with β > 0.6 and D < 150 mm the flange-tap tolerances cannot be relaxed.

Traps that are more common than a short run

  • Half-open upstream isolation valve — insufficiently developed flow, C drifts. Regulate through the downstream valve.
  • Elbow right in front of the plate — the straight run may be long enough, but the fitting type is wrong and the profile is still uneven. The standard's tables are per fitting combination; do not convert them by hand.
  • Missing or blocked taps — no Δp signal, or a polluted one.
  • A plate that is never cleaned — scaling or corrosion pushes the upstream surface roughness above 10⁻⁴·d and the C deviation exceeds the limit. Scaling and fouling mean shutting down for maintenance.
  • Using the plate in pulsating flow — explicitly outside the scope of the standard; fit a flow conditioner vessel or change the principle.

Sources: GB/T 2624.1-2006 (identical to ISO 5167-1), GB/T 2624.2-2006 (table 3, straight runs), JB/T 7252-2018.

Tell me the straight run and fittings at site and I will price the options — a conditioner or the integral orifice, whichever makes more sense.