SKCK SKCKHui Ma · Flow measurement & instrumentation

Home / Tools

Selection check tools

The equations are taken from the GB/T 2624 series (identical adoption of ISO 5167). This is a check: feed in the bore and pressure drop you already have and see whether it holds up. Each result comes with the applicable limits checked one by one — if a limit is crossed, the quotation and the drawing cannot go further.

Basis: GB/T 2624.1–.4-2006, GB/T 2624.5/.6-2024 (identical to ISO 5167 series), GB/T 21446-2008.

These are theoretical values from the standard equations, not a formal design. A real design has to account for machining tolerance, edge thickness, material, C-value test data and the installation conditions — that one comes with a calculation sheet and test report issued by us.

Duty data

ρ₁ and μ are the upstream duty values. For gas, the absolute pressure is required — otherwise ε and τ cannot be computed.

Results

Fill in the left-hand fields and press Calculate.

Applicability check

    The orifice C value uses the Reader-Harris/Gallagher equation; Venturi tubes, cone and wedge devices use the constant or segmented forms from the standard, and the Venturi nozzle uses C = 0.9858 − 0.196β4.5. Flange taps and D < 71.12 mm require a correction per the standard; this tool gives the base value.

    Basis

    Basis of calculation

    • GB/T 2624.1-2006 General principles and requirements (identical to ISO 5167-1)
    • GB/T 2624.2-2006 Orifice plates (Reader-Harris/Gallagher discharge coefficient)
    • GB/T 2624.3-2006 Nozzles and Venturi nozzles
    • GB/T 2624.4-2006 Venturi tubes
    • GB/T 2624.5-2024 Cone device, GB/T 2624.6-2024 Wedge device
    • JB/T 7252-2018 Integral orifice flow meter
    • GB/T 21446-2008 Distance between orifice plate face and tapping centre
    Limits of this tool: it supports engineering estimation and first-stage selection; it is not a verification certificate. Supply is governed by the technical agreement and drawings confirmed by both parties; custody-transfer metering requires real-flow calibration or an accredited laboratory.

    Unsure about a duty? Send the medium, temperature, pressure, bore, flow range and installation conditions to me — I will check them against the standard and prepare the selection sheet with the quotation.