SKCK SKCKHui Ma · Flow measurement & instrumentation

Home / Articles / EM flowmeter: DN and range

SKLD electromagnetic flowmeter

Electromagnetic flowmeter: choosing DN and range

Updated 2026-11-21 · about 4 minutes

The measurement itself is simple — Faraday induction makes the voltage proportional to mean velocity, so the meter measures volumetric flow regardless of temperature, pressure, density or viscosity, and needs one calibration. The difficult part is getting DN and range right.

Electromagnetic flowmeter: split installation, working principle and performance
Electromagnetic flowmeter and its working principle: the fluid in blue, a pair of measuring electrodes above and below the pipe, and the magnetic field produced by the coil around it.

The selection order

Selection does not start with "how big a bore do I need". It starts with the two parts that touch the medium. The SIKE SKLD order is 5-4-3-2-1:

  1. medium temperature + working pressure → liner (many duties are decided here);
  2. corrosiveness + particulates → electrode;
  3. process flow → DN → bore and pressure class;
  4. control system protocol → transmitter;
  5. installation environment → protection class and explosion-proof type.

Get the order wrong and you get the classic rework: material fixed before the flow is known, or a transmitter whose protocol does not match the DCS.

Using the 1 m/s table to pick DN

The usual basis is the flow corresponding to 1 m/s:

Q(m³/h) = 0.002827 · D² (D in mm)

DN100 corresponds to 28.3 m³/h, DN500 to 708 m³/h. Any other velocity is a simple multiple of that table. In practice you back-calculate the DN so that the usual flow sits in the middle of the recommended range — not at its upper end.

DNQ at 1 m/s (m³/h)Recommended range (m³/h)Max measurable (m³/h)
507.081.7 – 570.708 – 70.8
8018.14 – 1451.81 – 181
10028.37 – 2262.83 – 283
15063.715 – 5106.37 – 637
20011327 – 90411.3 – 1130
25017742 – 141617.7 – 1770
30025561 – 204025.5 – 2550
400453109 – 362445.3 – 4530
500708170 – 566470.8 – 7080
6001019245 – 8152101.9 – 10190
7001387333 – 11096138.7 – 13870

Source: appendix 1 and 2 of the SIKE SKLD manual. DN800 and above are proportional estimates; higher pressure classes are available on request.

Recommended range vs maximum measurable flow

The maximum measurable flow is a hardware limit. Running at it is possible but accuracy and stability suffer clearly. What you should use is the recommended range, near its middle, keeping about a third of the range in reserve.

  • usual flow in the middle of the recommended range → good range utilisation;
  • close to the lower end → careful: below 0.3 m/s viscosity effects grow;
  • close to the upper end → warning: above 10 m/s liner and electrode wear accelerates;
  • beyond the maximum measurable → wrong, the DN must be increased.

The SKLD turndown reaches 150:1, so small flows are not the problem — provided the velocity does not drop too low.

Set the range to the maximum flow

This is the most common commissioning error: the process says "we usually run at 40 m³/h", so the transmitter is set with 40 m³/h as full scale — and then alarms on over-range as soon as production steps up. The range must be set to the maximum process flow, with the usual operating flow written on the nameplate. Keep the two apart.

Liner and electrode: medium first, then temperature and pressure

MediumLinerElectrode
General acids / alkalisPTFEStainless steel
Slurry / coal slurryPUTungsten carbide
Seawater / chloridePTFETitanium
Strong acid (HCl)PTFE or FEPHastelloy (HB)
Hot strong acidFEPTantalum (Ta)
Steam (permeation)Reinforced PFAStainless steel
Neutral water (economy)NeopreneStainless steel

Source: SIKE SKLD liner/electrode selection matrix. Exact grades and temperature limits are confirmed in the technical agreement.

Three preconditions people forget

  • Conductivity ≥ 5 μS/cm. That is the floor of the principle. Ultra-pure water, distillate and fuels are not conductive — an EM meter simply cannot work there.
  • Straight run: 5 DN upstream, 2 DN downstream; the two electrodes must sit on the same horizontal line, which is regularly forgotten on vertical installations.
  • The signal cable shield must be earthed, otherwise interference arrives directly in the flow signal.

Most site alarms are not a failed instrument: the empty-pipe, excitation and over/under-range alarms cover most cases. Check full pipe, then the coils, then the range setting — seven out of ten problems are solved that way.

Sources: SIKE SKLD electromagnetic flowmeter manual (豫制 02000040, 2016F103-41), chapters 3–6; JB/T 9248-1999.

Put your maximum process flow into the EM flowmeter lookup, or send me the data and I will return DN, range and materials.